    题名: A reevaluation and discussion on the threshold limit for hemolysis in a turbulent shear flow
    作者: 盧博堅;Lu, Po-chien;Lai, Ho-cheng;劉佳興;Liu, Jia-Shing
    贡献者: 淡江大學水資源及環境工程學系
    日期: 2001-10-01
    上传时间: 2010-03-26 16:58:06 (UTC+8)
    出版者: Elsevier
    摘要: One major factor of red blood cell damage induced by artificial heart valves is the magnitude of turbulent shear stresses in the flow field. An often-cited threshold for hemolysis is 400 N/m2 (Sallam and Hwang, Biorheology 21 (1984) 783). This value, however, was measured with a one-component laser Doppler anemometer and was not calculated from the major principal Reynolds shear stress, therefore underestimatng the threshold. This study performed flow field measurements under similar conditions, using a two-component laser Doppler anemometer and determining major principal Reynolds shear stress, to reevaluate the hemolytic threshold as 800 N/m2 with an exposure time of 1 ms. The Kolmogorov length scales, approximately 9 μm, are capable of inflicting direct damage to the red blood cells. These results will serve as a more accurate reference in future heart valve design and testing.
    關聯: Journal of biomechanics 34(10), pp.1361-1364