    Title: Enzymatic modified glucose sensor with heterodyne interferometer
    Authors: Zhao, Jia-Hui;Hsu, Cheng-Chih;Wu, Chyan-Chyi;Lee, Ju-Yi;Dai, Ching-Liang
    Keywords: Glucose concentration;Heterodyne interferometer;Glucose oxidase;Enzyme;Glucose sensor
    Date: 2019-09
    Issue Date: 2020-03-28 12:10:34 (UTC+8)
    Abstract: In this study, an enzymatic modified cuvette was fabricated and adopt to measure the
    glucose concentration with heterodyne interferometer. Based on the optical configuration, the
    calibration of the modified cuvette was no longer required. The results indicated that the resolution
    of the proposed method was approximated of 3 mg/dl. The calibration curve showed high linearity
    within the dynamic range of 0 ~ 500 mg/dl.
    Relation: Proceedings of ISMTII 2019
    Appears in Collections:[Graduate Institute & Department of Mechanical and Electro-Mechanical Engineering] Proceeding

    Files in This Item:

    File SizeFormat
    index.html0KbHTML164View/Open

    All items in 機構典藏 are protected by copyright, with all rights reserved.


    DSpace Software Copyright © 2002-2004  MIT &  Hewlett-Packard  /   Enhanced by   NTU Library & TKU Library IR teams. Copyright ©   - Feedback