    題名: Multiple Silicon Nanowires with Enzymatic Modification for Measuring Glucose Concentration
    作者: Cheng-Chih, Hsu 1;Yu-Ching, Liao 1;Yen-Ting, Tsai 1;Hsin-I, Yeh 1;Wu, Chyan-Chyi
    關鍵詞: multiple nanowires;glucose sensor;enzymatic modification;top-down method
    日期: 2015-08-14
    上傳時間: 2017-02-24 02:11:25 (UTC+8)
    摘要: This study fabricated a multiple poly-Si nanowires sensor through a top-down method and immobilized glucose oxidase on the multiple nanowires for determining glucose concentration. The proposed sensor is 340 nm in width and uses five physically identical and parallel nanowires. The sensor contained nanowires of various lengths (3, 5, and 10 µm). Experimental results showed that sensor sensitivity is inversely proportional to nanowire length. The sensor with 3 µm in nanowire length exhibited a theoretical resolution of 0.003 mg/dL and the highest sensitivity of 0.03 µA/(mg/dL). Furthermore, the proposed sensor retains this performance when reused for up to 10 applications.
    關聯: Micromachines 6(8), pp.1135-1142
    DOI: 10.3390/mi6081135
    顯示於類別:[機械與機電工程學系暨研究所] 期刊論文